Bruker Avance 700 Reference and Temperature Calibration

Bar Break


Probe:
5 mm TXI (1H/13C/15N)
5 mm QXI (1H/13C/15N/31P)

Calculate using:
Actual temperature:
Set temperature:
K

Frequency values:
Hfreq :   MHz
Cfreq :   MHz
Nfreq :   MHz

Results:
Actual temperature: 298.66 K
Set temperature: 300.00 K
Hppm: 4.752
Cppm1: 56.017
Cppm2: 176.018
Nppm: 118.017
H DSSfreq: 700.1299699 MHz
Hfreq: 700.1332970 MHz
Cfreq1: 176.0572136 MHz
Cfreq2: 176.0783393 MHz
Nfreq: 70.9519249 MHz